AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 37
3.4.1.2.
X- Y axis Electrical Functions Diagram
The table below describes the electrical components of each PCB which effects the movement of the X-Y axis.
The arrangement is shown in the diagram.
DILAP21 PCB
Microprocessor and memory.
SIR4MOT PCB
Controls X - Y axis.
XGOLAV “X” PCB
Controls opto with which the encoder controls axis position.
INTEROP2 ”Y” PCB
Opto couplers
The Opto couplers, together with the selectors, control:
1) Y probe position
2) Z1 and Z2 probes position
3) ZZ probe position
INTERFA1 PCB
Connectors for wiring.
INTERFA2 PCB
Connectors for wiring.
AP22MOT PCB
Connectors for wiring.
X axis motor
Step motor
Y axis motor
Step motor
DILAP21 PCB
SIR4MOT
PCB
M
M
INTEROP2 “Y”
PCB
XGOLAV “X”
PCB
INTERFA1
PCB
AP22MOT
PCB
INTERFA2
PCB
X axis motor
Y axis motor
Memory
X axis
motor
driver
Y axis
motor
driver
Microprocessor