AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 32
3.3.3.2.
Substituting valve mechanism
a) Turn on instrument. Run XCalib Program (See Section 4). From “Dilutor” pulldown menu select “Calibrate
syringe” and follow instructions.
b) Click on “Remove syringe” to fully open syringe plunger.
c) Turn off machine
d) Remove AWG 15 tubing from valve joint
e) Unscrew syringe from valve mechanism in an anti-clockwise direction and extract it
f)
Remove tubing joined to valve mechanism
g) Remove valve connector by undoing fixing screw
h) Remove valve mechanism by undoing 2 fixing nuts
i)
Insert new valve mechanism and re-block it with 2 fixing nuts.
j)
Replace valve and connector
k) Replace hydraulic connections
l)
Replace syringe in movement shaft and screw it back into valve mechanism in a clockwise direction.
m) Turn on instrument again and run a prime to rinse hydraulic circuit
3.3.3.3.
Substituting 3-way Electro valve
Note: to replace electro valve follow procedure described in para. 3.11.2.1.
3.3.3.4.
Substituting 3-way valve mechanism
Note: to replace valve mechanism follow procedure described in para.3.11.2.1.
Valve mechanism
tubing join
Connector/
Electro-valve
fixing screw
Hexagonal head
screws