AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 83
4.7.7. Turn Light On
From “
Main functions
” pull down menu select “
Turn light on
” to check that internal
lights are functioning correctly for reading phase
If light does not come on check (and possibly substitute) the following - in order:
1) connections are correct
2) 2) Colight PCB.
3) 3) LED light bar (see ref. para. 3.7.4.4.).
4.7.8. Close Function
From “
Main functions
” pull down menu select “
Close function
” to close active
functions and return to initial display.
4.7.9. Exit
From “
Main Functions
” pull down menu select “
Exit
” to leave XCALIB programme.