AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 75
4.5. SETTINGS
4.5.1. HARDWARE CONFIGURATION
The Xcalib requests a hardware configuration of the instrument which is going to be used to be input.
All calibration procedures are disactived until the configuration is successfully loaded.
From “
Settings
” menu select “
Hardware configuration
” or directly click on
icon on main instrument bar.
The following display will appear:
This display can be used to set-up the following:
which path the XML folder will be associated with this configuration (which can be chosen from a list by
clicking Browse button);
which instrument configuration can be calibrated (from the pull-down menu);
which type of configuration will be associated with the device to be calibrated (from the pull-down menu).
To store the configuration settings click on “
Store
” and the programme displays a confirmation message for the User
as shown here: