AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 91
4.8.3. Calibrate LT
From “
Dilutor
” pull down menu select “
Calibrate LT
” .
The following display will appear:
Check and Calibrate LT liquid sensor
1. Ensure that sample rack heights have been correctly calibrated.
2. Prepare test solution: distilled water and dilutor liquid in 1/100 ratio
3. Insert test-tube with 200µl liquid sensor solution anywhere in sample rack.
4. Remove probe 1 tube.
5. From “
Sample number
” box type in or pull down value of position where test-tube with liquid is inserted
6. Select “
Measure LT
” command
7. The probe will automatically go down 10 steps
inside the test-tube to measure and evaluate LT settings
8. When measurements are completed, a report with all 10 results is displayed.
9. Reinsert tube in Probe 1