Scanning Electron Microscope
EM-30PLUS Operating Manual
94
5.7.2 BSE Setting
See [BSE Setting in 2.7.1 Setup].
5.7.3 Directions for BSE Mode
Preparations before and after selection of BSE Mode
1) The Signal of BSE varies rapidly depending on WD. Be sure to adjust the height of Stub
and set it to the optimum height. (recommended WD: 7~10mm)
2) When using BSE Mode, there can be a big difference in Spot Size from SE Mode.
Be sure to have the optimum electron beam come down by readjusting Wobble and
Gun Alignment.
3) Execute the optimization of Wobble in SE Mode in the Spot Size that you want to
measure in BSE Mode. (See [6.2.1 Wobble operation 1])
Method for selection of SE and BSE Mode
You can observe SE images when it is indicated in SE Mode, and BSE images when it is
indicated in BSE Mode. (Be sure to move to the state of BSE Detector ON for
observation.)
Optimization of BSE image
1) Select the appropriate Spot Size on the Set UP screen.
2) Execute the optimization on the Image Viewer Setting screen on the bottom right.