5.7 BSE Mode
BSE Mode refers to the image detected with Back Scattered Electrons(BSE) generated by
electron beam using BSE sensor.
BSE Signal is expressed in different brightness depending on the element components of
materials. If the material components are different, images can be detected as there is a
clear difference in the BSE images compared to the SE images.
SE Image
BSE Image
Removable BSE of EM-30PLUS equipment can be rotated manually and converted to the
bottom of electron beam or Stage entrance.
BSE Detector ON
BSE Detector OFF
When using BSE Mode, be sure to move BSE sensor to the bottom of electron beam.
When analyzing EDS, rotate Detector towards the Stage entrance and convert it for the
smooth detection of EDS Signal.
Method for manual rotation of BSE
1) Pull the body that has bolts in the vertical direction.
2) Rotate to the direction you want(bottom of electron beam or Stage entrance).
3) Place BSE in the fixed Hole in the upper body of BSE.
*Rotation must be executed by the person who completed training on rotation only.
Be sure not to touch the sensor to prevent contamination of BSE.