REM Alarm Indicator
2-8
Force FX Electrosurgical Generator C Service Manual
REM Alarm Indicator
This indicator illuminates red until you properly apply a REM Polyhesive
TM
Patient Return
Electrode to the patient and connect it to the generator. Then the indicator illuminates
green. (When you connect an electrode without the REM safety feature, the indicator
does not illuminate.)
If the REM system senses an alarm condition, the indicator flashes red until you correct
the alarm condition—then the indicator illuminates green. (If you are using a return
electrode without the REM safety feature, the red indicator light is extinguished when
you correct the alarm condition.)
Rear Panel
Controls and receptacles on the rear panel
Volume control
Power entry module
Monopolar
Footswitch
receptacles
Option panel
Bipolar
Footswitch
receptacle
Equipotential grounding lug
Use to connect the generator to
earth ground.
Summary of Contents for Force FX
Page 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Page 30: ......
Page 180: ......
Page 210: ......
Page 246: ......