Responding to System Alarms
6-12
Force FX Electrosurgical Generator C Service Manual
10
Software malfunction. Invalid
alarm number (number shown
on terminal using serial port)
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
11
Internal diagnostics. Invalid
activated power, function,
mode, or key request mode
echoed by feedback
microcontroller
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
12
Diagnostics/microcontroller
malfunction. T_ON_ERR test
failed
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
13
Main microcontroller unable
to access GEN_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
14
Internal diagnostics.
Calibration data checksum
error on main microcontroller
(data in error shown on
terminal using serial port)
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
16
Diagnostics/microcontroller
malfunction. T_ON average
test failed
Calibrate the ECON factor. Refer to
.
If the alarm number reappears, refer to
68
Main microcontroller unable
to access GEN_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
69
70
71
80
Software malfunction.
flash_power_state value
outside state range
Software malfunction.
key_req_mode value outside
state range
Software malfunction.
1 value outside state
range
Software malfunction.
*input[i].p_state value outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
Number
Description
Recommended Action