Front Panel Replacement
7-10
Force FX Electrosurgical Generator C Service Manual
c.
Disconnect the membrane display ribbon cable from the Display board.
d.
Disconnect the grounding clips from each side of the bezel by sliding the clips
away from the bezel until they are loose and held only by the Display board.
2.
Remove the REM module and reinstall it on the new front panel.
a.
Remove the two screws that secure the REM module to the front panel. Then,
remove the bracket that covers the REM module and remove the REM module.
b.
From inside the new front panel, insert the REM module into the front panel
opening.
c.
Place the bracket over the REM module and install the two screws that secure the
module to the new front panel.
3.
Remove the power switch and reinstall it on the new front panel.
a.
Inside the front panel, press the four tabs that secure the power switch and push
the switch (and connected cable) out the opening.
b.
Outside the new front panel, route the power switch cable through the power
switch opening.
c.
Position the power switch with the On (
|
) switch above the Off (
O
) switch. Press to
snap the switch into place.
4.
Install the Display board on the new front panel.
a.
Connect the membrane keyboard ribbon cable to the Display board.
b.
Connect the CEM switch cable to the Display board.
c.
Connect the grounding clips to each side of the bezel by sliding the clips toward
the bezel until they are snug.
d.
Position the Display board over the front panel posts and install the four screws
that secure the board to the front panel.
Install the Front Panel Assembly
1.
Position the front panel assembly in front of the chassis and carefully slide it into the
electrical contacts on the Power Supply/RF board.
2.
Install the four screws that secure the front panel to the chassis.
3.
Connect the REM wires to the Power Supply/RF board.
a.
Connect the white wires with the 2-pin nonlocking connector to J15.
b.
Connect the red wires with the 2-pin locking connector to J17.
4.
Connect the power switch cable to J10 on the Power Supply/RF board.
5.
Connect the Display board ribbon cable to the Control board and lock the connector.
6.
Position the cover above the chassis and slide it down. Install the five screws that
secure the cover to the chassis.