Output Power vs. Resistance Graphs
3-30
Force FX Electrosurgical Generator C Service Manual
Standard Bipolar Mode
Output power vs. impedance for Standard bipolar 35W mode
Output power vs. impedance for Standard bipolar 70W mode
Note:
In the Bipolar Standard mode only, the power curve will reflect both High and Low
output values between the impedance range of 475-600 ohms. This is based on the
sensed impedance value the generator calculates, and the tolerances of the sensing
circuits. This is a function of the generator, as the software is designed to reduce (switch
to Low output) the output power when the impedance exceeds a reference
Load Resistance (ohms)
Output Power (watts)
Load Resistance (ohms)
Output Power (watts)