Low Voltage Power Supply Replacement
7-22
Force FX Electrosurgical Generator C Service Manual
Low Voltage Power Supply Replacement
Equipment required
• Phillips screwdriver
Remove the Low Voltage Power Supply
1.
Turn off the generator. Disconnect the power cord from the wall receptacle.
2.
Remove the five screws that secure the cover to the chassis. Lift the cover off the
chassis. Set the cover and screws aside for reinstallation.
3.
Remove the three screws (one plastic and two metal) that secure the support bracket
to the heat sinks. Save the screws and bracket for reinstallation.
4.
Disconnect and remove the low voltage power supply.
a.
Disconnect the Power Supply/RF board cables from the low voltage power supply
(red spade lug and 3-pin connector at the rear; 6-pin connector at the front).
Mark the cables for reinstallation.
b.
Remove the four screws that secure the low voltage power supply to the Power
Supply/RF board. Save them for reinstallation.
c.
Lift the low voltage power supply off the Power Supply/RF board.
Install the Low Voltage Power Supply
1.
Place the new low voltage power supply on the Power Supply/RF board and install the
four screws that secure it to the board.
2.
Facing the front of the unit, connect the cables from the Power Supply/RF board to
the new low voltage power supply.
Warning
Electric Shock Hazard
To allow stored energy to dissipate after power is disconnected, wait at
least five minutes before replacing parts.
Precaution
The generator contains electrostatic-sensitive components. When repairing the generator, work at
a static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized generator. Handle circuit boards by their
nonconductive edges. Use an antistatic container for transport of electrostatic-sensitive
components and circuit boards.