Setting Up the Generator
Setup, T
ests, and Adjustment
Force FX Electrosurgical Generator C Service Manual
5-3
• Activation tones sound to verify that the speaker is working properly.
7.
If the self-test is successful, a tone sounds. Verify the following:
• Indicators above the default mode buttons (Standard bipolar, Pure cut, and
Fulgurate coag) illuminate green.
• Each display shows a power setting of one watt.
• The REM Alarm indicator illuminates red.
If the self-test is not successful, an alarm tone sounds. A number may momentarily
appear in the Cut display and, in most cases, the generator is disabled. Note the
number and refer to
Important
If the coag mode has been optionally changed to default to Desiccate or Spray, that
corresponding indicator illuminates after the self-test is performed successfully.