Responding to System Alarms
Troubleshooting
Force FX Electrosurgical Generator C Service Manual
6-11
Responding to System Alarms
When a system alarm condition exists, an alarm tone sounds and a number flashes in the
Cut display. The generator is disabled until the condition is cleared.
Most system alarms require some action on your part to correct the condition; however,
some are corrected automatically. Use the following table to determine how to correct an
alarm condition.
After correcting the alarm condition, verify that the generator completes the self-test as
described in Chapter 5,
.
Number
Description
Recommended Action
0
1
Main microcontroller failed to
hold feedback microcontroller
in reset
Main microcontroller failed to
power up feedback
microcontroller
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
2
3
Feedback microcontroller
failed to power up and
initialize RAM in time allotted
Feedback microcontrolled
failed to checksum battery-
backed RAM data in dual-port
RAM in time allotted
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
4
Internal diagnostics
Main microcontroller CPU test
failed
Replace the Control board. Refer to
Calibrate the generator. Refer to
If the alarm number reappears, record the
number and call the Covidien Service Center.
5
6
7
Main microcontroller unable
to access FEEDBACK_SEM
semaphore
Main microcontroller unable
to access ECON_SEM
semaphore
Main microcontroller unable
to access KEY_ACTIVE_SEM
semaphore
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to