Responding to System Alarms
6-14
Force FX Electrosurgical Generator C Service Manual
Number
Description
Recommended Action
112
113
114
115
116
117
118
119
Software malfunction.
cal_mode value in
ca_clock_date_setup( ) outside
state range
Software malfunction.
cal_state value in
ca_clock_time_setup( ) outside
state range
Software malfunction.
cal_mode value in
ca_clock_time_setup( ) outside
state range
Software malfunction.
cal_state value in
ca_coag_leakage_cal( )
outside state range
Software malfunction.
cal_state value in ca_econ_cal(
) outside state range
Software malfunction.
econ_cal_state value outside
state range
Software malfunction.
cal_state value in ca_rem_cal( )
outside state range
Software malfunction.
rem_cal_state value outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
120
Calibration malfunction.
Calibration value(s) outside
acceptable range
Repeat the failing calibration step. If the alarm
number reappears, record the number and call
the Covidien Service Center.
121
Software malfunction.
cal_mode value in
ca_generator_setup( ) outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
122
Open circuit REM sense failure
Repeat the failing calibration step. If the alarm
number reappears, record the number and call
the Covidien Service Center.
Summary of Contents for Force FX
Page 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Page 30: ......
Page 180: ......
Page 210: ......
Page 246: ......