Monopolar Cut Controls
Contr
ols, Indicators,
and Receptacles
Force FX Electrosurgical Generator C Service Manual
2-5
Monopolar Cut Controls
Buttons and indicators for cut controls (monopolar)
Cut indicator
When you activate cut, this
indicator illuminates yellow and
an activation tone sounds.
Power buttons
Press
to increase the power.
Press
to decrease the power.
Mode indicators
Illuminate green when
you press the
corresponding mode
button.
Blend mode button
Select for slower cutting and
additional hemostasis.
Pure mode button
Select for an even cut with little or no
hemostasis. This is the default monopolar cut
mode.
Low mode button
Select for a cut with little
or no sparking.
Cut display
Shows the power
setting, in watts, for the
selected mode.
Summary of Contents for Force FX
Page 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Page 30: ......
Page 180: ......
Page 210: ......
Page 246: ......