![background image](http://html1.mh-extra.com/html/covidien/force-fx/force-fx_service-manual_2673882172.webp)
Responding to System Alarms
6-18
Force FX Electrosurgical Generator C Service Manual
200
201
Internal diagnostics. Main
microcontroller SEML line may
be stuck
Internal diagnostics. Feedback
microcontroller SEMR line may
be stuck
1.
Replace the Control board. Refer to
2.
Calibrate the generator. Refer to Calibrating
the Generator
If the alarm number reappears, record the
number and call the Covidien Service Center.
202
Internal diagnostics. Main
microcontroller watchdog test
failed
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
203
206
207
Internal diagnostics. Feedback
microcontroller watchdog test
failed
Software malfunction.
doserr_test_state value in
st_m_doserr_test outside state
range
Software malfunction.
doserr_test_state value in
st_fb_doserr_test outside state
range
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
208
209
Main microcontroller timer
interrupt failed
Feedback microcontroller
timer interrupt failed
1.
Replace the Control board. Refer to
2.
Calibrate the generator. Refer to Calibrating
the Generator
If the alarm number reappears, record the
number and call the Covidien Service Center.
210
211
Software malfunction.
data_type value outside state
range
Software malfunction.
which_data value outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center
212
Generator not calibrated
Refer to
213
Internal diagnostics. Firmware
not compatible with hardware
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
Number
Description
Recommended Action
Summary of Contents for Force FX
Page 1: ...Service Manual Force FX TM Electrosurgical Generator C...
Page 30: ......
Page 180: ......
Page 210: ......
Page 246: ......