Using the RS-232 Serial Port
Setup, T
ests, and Adjustment
Force FX Electrosurgical Generator C Service Manual
5-43
keying ENABLED
When activation ceases, the following message appears:
key request removed
Enter the Commands
You can enter commands to change settings or display data about the generator. Each
command should begin with a colon (:) and end with Enter. The # symbol represents a
numeric value that you enter with specific commands as noted.
Data Displays
Some commands result in the display of data for each mode, including CEM. The display
format is as follows:
In data displays, numbers represent the Force FX
Electrosurgical Generator C modes. You
also use these numbers when you enter commands to select a specific mode.
Bipolar Modes
0 Precise
1 Standard
2 Macro
Cut Modes
0 Low
1 Pure
2 Blend
Coag Modes
0 Desiccate
1 Fulgurate
2 Spray
Bipolar Precise
Bipolar Standard
Bipolar Macro
Default
Fulgurate
Cut Low
Cut Pure
Cut Blend
Cut CEM
Coag Desiccate
LCF Fulgurate
Coag Spray
Coag CEM