Responding to System Alarms
Troubleshooting
Force FX Electrosurgical Generator C Service Manual
6-13
81
Internal component
malfunction. Optoisolator test
failed
Do not attempt to use the generator. Record the
number and call the Covidien Service Center.
90
95
Generator model number in
main microcontroller Flash
Memory not Force FX
Electrosurgical Generator C
model number
Generator model number in
main microcontroller Flash
Memory not Force FX
Electrosurgical Generator C
model number
Turn the power switch off (O) then on (I) again.
If error reappears, replace the Control board.
Refer to
100
101
102
103
104
105
110
111
Software malfunction.
rem_update_state value
outside state range
Software malfunction.
rem_pad_state value outside
state range
Software malfunction.
rem_state value in
re_rem_control( ) outside state
range
Software malfunction.
rem_state value in
re_non_rem_control( ) outside
state range
Software malfunction.
rem_flash_state value outside
state range
Software malfunction.
rem_led_state value outside
state range
Software malfunction.
cal_state value in
ca_generator_setup( ) outside
state range
Software malfunction.
cal_state value in
ca_clock_date_setup( ) outside
state range
Turn off, then turn on the generator. If the alarm
number reappears, record the number and call
the Covidien Service Center.
Number