Using the RS-232 Serial Port
5-44
Force FX Electrosurgical Generator C Service Manual
Commands for Selecting Modes and Power Settings
Command
Action
:FBM#
Specify the bipolar mode (for #, enter the number for the
desired mode):
0 = Precise
1 = Standard
2 = Macro
:FBP#
Specify the bipolar power setting (# = 1, 2, 3, … 65, 70).
:FBU
Increment the bipolar power setting to the next higher value.
:FBD
Decrement the bipolar power setting to the next lower value.
:FTM#
Specify the cut mode (for #, enter the number for the desired
mode):
0 = Low
1 = Pure
2 = Blend
:FTP#
Specify the cut power setting(# = 1, 2, 3, … 290, 300).
:FTU
Increment the cut power setting to the next higher value.
:FTD
Decrement the cut power setting to the next lower value.
:FGM#
Specify the coag mode (for #, enter the number for the desired
mode):
0 = Desiccate
1 = Fulgurate
2 = Spray
:FGP#
Specify the coag power setting (# = 1, 2, 3, … 110, 120).
:FGU
Increment the coag power setting to the next higher value.
:FGD
Decrement the coag power setting to the next lower value.
:FR
Recall the previous front panel settings.