Display Board Replacement
7-6
Force FX Electrosurgical Generator C Service Manual
d.
Disconnect the grounding clips from each side of the bezel by sliding the clips
away from the bezel until they are loose and held only by the Display board.
Install the Display Board
1.
Connect and attach the Display board to the front panel.
a.
Connect the membrane keyboard ribbon cable to the Display board.
b.
Connect the CEM switch cable to the Display board.
c.
Connect the grounding clips to each side of the bezel by sliding the clips toward
the bezel until they are snug.
d.
Position the Display board over the front panel posts and install the four screws
that secure the board to the front panel.
2.
Install the front panel assembly.
a.
Position the front panel assembly in front of the chassis and carefully slide it into
the electrical contacts on the Power Supply/RF board.
b.
Install the four screws that secure the front panel to the chassis.
c.
Connect the power switch cable to J10 on the Power Supply/RF board.
3.
Connect the REM wires to the Power Supply/RF board.
a.
Connect the white wires with the 2-pin nonlocking connector to J15.
b.
Connect the red wires with the 2-pin locking connector to J17.
4.
Install the Control board.
a.
Position the Control board over the Power Supply/RF board with the Control
board components facing the rear panel. Fit the edges of the board into the slots
on the heat sinks. Slide the board down, carefully fitting the edge connector into
the matching connector on the Power Supply/RF board.
b.
Connect the Display board ribbon cable to the Control board and lock the
connector.
5.
Position the cover above the chassis and slide it down. Install the five screws that
secure the cover to the chassis.
Summary of Contents for Force FX