Confirming Outputs
Setup, T
ests, and Adjustment
Force FX Electrosurgical Generator C Service Manual
5-21
b.
Release the footswitch pedal.
c.
Based on the voltmeter setting and the current transformer you are using,
calculate and record the output current.
6.
Press the Low cut button and repeat step 5 (above).
7.
Press the Blend button and repeat step 5 (above).
8.
Verify that the generator output for each mode is 499 ± 38 mA
rms
.
If the output is outside the specified range, calibrate the monopolar output as
described in
.
Then repeat this procedure. If
the output for one or more cut modes remains outside the specified range, call the
Covidien Service Center
.