Standards and IEC Classifications
3-18
Force FX Electrosurgical Generator C Service Manual
Cables Used for the Force FX EMC Test
Force FX EMC Test—Cables Used
Model
Function
Length
Location
207002060
AC cable
15 ft.
Rear panel
Multicontact Part Number:
15.2013-46020
Equipotential cable
15 ft.
Rear panel
E2450HDBNSB
Two-button pencil
15 ft.
Front panel mono 1
E2450HDBNSB
Two-button pencil
15 ft.
Front panel mono 2
E7507DB
Adult REM pad
15 ft.
Front panel REM
E0020V / E4073CT
Foot switching bipolar
cord / bayonet forceps
15 ft.
Front panel bipolar
E6019
Dome bipolar footswitch
15 ft.
Rear panel bipolar
E6008B
Monopolar footswitch
15 ft.
Rear panel mono 1
E6008B
Monopolar footswitch
15 ft.
Rear panel mono 2
VLCM
Bipolar current monitor
24 ft.
Rear panel
EA3730
EKG blanking cable
5 ft.
Rear panel
Warning:
The use of accessories, other than specified in the Force FX-C User’s Guide and Service
Manual, may result in increased emissions or decreased immunity of the Force FX-C generator.
Only the use of Covidien-branded monopolar cables and accessories, bipolar cables and
accessories, and footswitches are approved for use with the Force FX generator. Use of non-
Covidien-branded cables and accessories have not been assessed and therefore do not comply
with the safety and EMC requirements of IEC 60601-1, IEC 60601-1-2, and/ or IEC 60601-2-2.
Summary of Contents for Force FX