2201R Mobile Tester SCPI Reference Guide
Chapter 4 Common commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
43
Introduction
The common commands are understood by all SCPI and IEEE 488.2 instruments. Their purpose is to perform general
tasks and to read or work with the registers common to all SCPI instruments. The following common commands have
been implemented.
General common commands
:REBoot
Syntax
:REBoot
Parameters
There are no parameters.
Description
Reboots the operating system. The current settings are not affected by this command.
Query
There is no query form of this command available.
Example
:REBoot
*CAL?
Syntax
*CAL?
Parameters
There are no parameters.
Description
The command form is not available.
Query
Returns date of last calibration by Cobham in the form yyyy,mm,dd
Example
*CAL?
Returns "2014,12,31"