* Relationship between the load cell output and input sensitivity.
Example 1) Number of load cell: 4 ea
Rated capacity of load cell: 500 Kg
Rated output of load cell: 2mV/V
Value of a division: 0.10 Kg
Applied voltage of load cell: 5V (= 5,000 mV)
According to the equation
(5000 mV* 2 mV * 0.1 Kg)/(500 Kg * 4) = 0.5 ≥ 0.2 uV
As the calculated value is greater than 0.2 uV, this weight system design has no problem.
Note 3. It can check the mV/V value in the testmode3
Applied voltage of load cell x Output voltage of load cell x Value of a division
0.2 uV ≤ ------------------------------------------------------------------------------------------------
Rated capacity of load cell x Number of load cell