![CAS CI-400 Series Owner'S Manual Download Page 8](http://html1.mh-extra.com/html/cas/ci-400-series/ci-400-series_owners-manual_2565928008.webp)
9
1-3. Analog and A/D Conversion
Input sensitivity
0.3 uV/D(Certification), 0.2 uV/D(None)
Zeroing range
0 ~ 2 mV/V
Loadcell supply voltage
DC 5V
Maximum signal input voltage
3.0mV/V
Zero variation with temperature
ZERO : ±10 PPM /
℃
SPAN : ±10 PPM /
℃
Input Noise
±0.6
㎶
P.P
Input Impedance
10
㏁
A/D conversion method
∆∑
A/D resolution
1,000,000 Count(20bit)
A/D conversion speed
350 rounds / Sec
Non-straightness
0.01 % FS
1-4. Digital and Display
Span Calibration
Full Digital Calibration (Single automatic span Calibration)
Display
7-Segment 3 Digits Red FND 2EA
(Font : 11 (W) X 20(H) mm)
7-Segment 3 Digits Red FND 8EA (CI-405,407)
(Font : 4 (W) *8 (H) mm)
Key
Number Key and Function Key (15 EA)
Division
X1, X2, X5, X10, X20, X50
Display Below Zero
“-”minus
Tare Subtraction
Full capacity
Sign for status
Stable, Zero, Tare, Gross/Net, Auto, Print, Hold, RX, TX (CI-401)
Stable, Zero, Tare, Gross/Net, SP1, SP2, SP3, SP4, Comm
(CI-405, CI-407)