117
11-2 Errors that can occur in weighing mode
Error
Cause
Solution
Err 01
Due to unstable load the weighing scale cannot
be reset.
Move the weighing scale to a flat, vibration-free
surface and turn on the power.
Err 02
Load cell is not properly connected or there is a
problem with the A/D convertor.
Check if the main body and the load plate are
connected properly.
Err 08
The zero-set key, container key and start key are
set not to function when the load is unstable.
In F22 of the Conversion Mode, set the zero-set key,
container key and start key activation conditions
based on the environment of use.
Err 09
The current weight exceeds the zero-set range.
In F09 of Conversion Mode set the activation range
to within 2% or 10% of the maximum capacity.
Err 10
The weight of the container being designated
exceeds the maximum capacity of the weighing
scale.
Set the container weight to be less than the maximum
capacity.
Err 13
When calibrating the weight the set zero-set has
been exceeded.
Check the status of the load plate and redo the weight
calibrations.
Err 15
In Command Mode, the Item Code has
exceeded the set range.
Check the Item Code range
Err 82
There is a problem with the A/D convertor.
Please contact the A/S center.
OVER
The current load on the load plate is too heavy,
and exceeds the allowance limit.
Do not place weight exceeding the capacity limit on
the weighing scale. If the load cell is damaged, it
must be replaced.
Summary of Contents for CI-400 Series
Page 1: ... 2 ...
Page 10: ...11 2 Specifications in Appearance 2 1 External Dimension ...
Page 29: ... 30 4 2 How to Seal the Indicator Sealing ...
Page 93: ... 94 Connection of Input Output Input AContact Output AContact ...
Page 117: ... 118 MEMO ...
Page 118: ...119 MEMO ...
Page 119: ... 120 MEMO ...
Page 120: ...121 ...