CAL 8(Gravity correction)
Function: Gravity Adjustment
Used key
Display
Descriptions
Using numeric keys
~
Enter
an initial gravity value.
= Set,
=Cancel
Push
when entering a decimal
G-CAL
It means you accessed to the
menu for the gravity adjustment.
Gr-CAL
9.XXXX
Set the gravity for the production
place.
Gr-SET
9.XXXX
Set the gravity for the place to
use the product.
Reference 1. If the gravity between the indicator production location and the location of use is different, gravity can be
corrected using this function.
CAL 9(Dual range setting)
CAL 9-1
Function: Setting Dual Range
Range of set value: 0 ~ 1
Used key
Display
Descriptions
Set the use status for dual function
= Set,
=Cancel
DUAL- 0
Dual range function is not used.
DUAL- 1
Dual range function is used.
Reference 1. If the resolving power is greater than 1/10,000, "OVER" will be displayed and will