CI-407 Relay Timing graph
<Limit mode
>
Relay Operation Graph upon Setting (F6.03- 0)
1. Required set value input : Final(SP7) > Step 1(SP1) > Step 2(SP2) > Fall(SP3)
2.
SP1:
Step1,
SP2:
Step 2,
SP3:
Fall,
SP5:
Upper Limit NG,
SP6:
Lowest Limit NG,
SP7:
Final
2. Near zero output is according to the specified range in F1-23
3. Relay Output
Step 1 : ON when {(SP7 – SP1 ) ≤
Weight}
Step 2 : ON when {(SP7 – SP2 ) ≤
Weight}
Step 3 : ON when {(SP7 – Fall ) ≤
Weight}
Lowest Limit NG : ON when {
Weight
< (SP7 – SP6)}
Upper Limit NG : ON when {(SP7+ SP5) <
Weight}
Near Zero : F1-23 set value ≥
Weight
≥ 0 range output
4. Status lamp in the front panel is operated in the same manner as the RELAY output.