AFX SERIES® OPERATION MANUAL
SECTION 10:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 352 of 402
10.9.1
Unit Settings
Unit settings determine the mode of operation of the power source. This includes phase mode,
voltage range, output mode, update phase angle and ramp time for any output value changes
made.
Output enable at power on allows the unit to power up with the output enabled.
The Enable Current Overload setting allows short duration overloads up to 30% over the
normal continuous mode current limit set point.
Phase Rotation selection changes output phasing between positive (leading) and
negative (lagging) phase rotation in three phase output mode.
The maximum adjustment limit for continuous self-calibration mode and fault
generation on saturation can be set from this screen as well.
The right hand side of the screen contains Transient mode execution settings.
Language selections are either ENGLIGH or CHINESE.
These can all be set from the CONFIGURATION -> UNIT SETTINGS screen shown below.