AFX SERIES® OPERATION MANUAL
SECTION 7:
Rear Panel, Connectors and Protection
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 178 of 402
Transient start mode:
SYSTem:DIO:STROBE:TRANsient 0/1
PROGram:TRANsient:TRIGger:OUTput 0/1
(alias)
Output Relay mode:
SYSTem:DIO:STROBE:OUTPutstate 0/1
Program Change mode:
SYSTem:DIO:STROBE:SOURce 0/1
The following specifications apply to the external trigger output:
Output Voltage @ 0.4 mA Logic Low Vout < 0.4 V
Log High Vout > 4.6 V
Max. Current
± 10 mA
Output Impedance
100
Pulse Width
190 us ± 10 us
For Transient Trigger mode
2.0 msec ± 10
s
For OUTPutstate & SOURce
modes
Figure 7-7: Transient Trigger Output Pulse
Figure 7-8: Function Strobe Output Pulse