AFX SERIES® OPERATION MANUAL
SECTION 6:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 106 of 402
6.4.8
PROGRAM Soft Keys
The following five soft keys are available on the main PROGRAM screen.
SOFT KEY
Description
Presets (SK1)
Changes all soft keys to up to five preset values that may be used to
change the selected parameter to any of the preset values shown on
the soft key labels. This provides a quick way to set commonly used
values. On AFX models with firmware revision 2.0.0 or higher, it is
also possible to reprogram these presets to user specific values. Refer
to Section 6.4.5, “Customizing Output Programming Preset Soft Key
Values” on page 104 for details.
The following presets are offered for the parameters listed:
Frequency
50Hz, 60Hz, 400Hz, 800Hz, 1200Hz
Voltage AC
115V, 230V, 300V
Voltage DC
MAX, 200V, 0V, -200V, MIN
Current limit
MAX, 15A, 4A, PEAK CURRENT
Power limit
MAX, 4kW, 2kW, 1kW
kVA limit
MAX, 4kVA, 2kVA, 1kVA
Protection
(SK2)
Displays the POWER AND CURRENT PROTECTIONS screen shown
below.
There are two check box options and one data field on this screen
that can be set as needed:
Enable RMS current protection
Activates the programmable
current limit function at the value
set in the main program screen
Enable power protection
Activates power protection at W
and kVA set points.
Trip time:
Determines how long the power
source will allow the power level
setting to be exceeded before
tripping off.