Command Syntax
*RST
Description
RESET. The *RST command (reset) has the same effect as an IEEE-488
Device Clear bus command but can be used over the RS232C, USB or
LAN interface as well. This command resets the unit to its power on
default state.
Parameters
None
Parameter Format
n/a
RESET STATE
FORM
3
VOLT:MODE
AC
VOLT:AC
0.0000
COUPLING
DC
VOLT:DC
0.0000
RANGE
AC
CURR:AC
41.6667
CURR:DC
20.8333
Command Syntax
*SRE <nr1>
Description
Before reading a status register, bits must be enabled. This command
enables bits in the service request register. The current setting is
saved in non-volatile memory.
Parameters
0-255
Parameter Format
<nr1>
Example
*SRE 255
Query Format
*SRE?
Description
Reads the current state of the service request enable register. The
register is cleared after reading it. Refer to section 8.12 for register
bit values.
Returned Data Format
<nr1>
Query Example
*SRE?
255
Query Format
*STB?
Description
Status Byte Query. The *STB? query returns the contents of the
status byte register (STB). After this query, the content of the STB
register is reset. Refer to section 8.12 for register bit values.
Returned Data Format
<nr1>
Query Example
*STB?
4
Command Syntax
*TRG
Description
Triggers pending operation.
Parameters
None
Parameter Format
n/a
Command Syntax
*WAI
Description
Prohibits the instrument from executing any new commands until all
pending overlapped commands have been completed.
Parameters
None
Parameter Format
n/a