AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 214 of 402
Query Syntax
OUTPut:DISABLEPHase:MINimum
OUTPut:DISABLEPHase:MAXimum
OUTPut:DISABLEPHase:DEFault
Description
These query commands return minimum, maximum and default
Output Disable Phase values respectively. Available range is -0.01
(Random) through 360.0.
Note: Available in units with Firmware revision 1.6.6 or higher.
Returned Data Format
<nr2>
Query Example
OUTP:DISABLEPH:MIN?
-0.010
OUTP:DISABLEPH:MAX?
360.000
OUTP:DISABLEPH:DEF?
0.000
Command Syntax
OUTPut:ZEROprogram
Description
This command sets the programmed output voltage to zero before
opening the output relay when the OUTP OFF | 0 command is sent.
This feature is disabled by default but can be enabled be sending
OUTP:ZERO ON | 1.
In this mode, the output of the power source goes to a low
impedance state before disconnecting the load.
Note: Available in units with Firmware revision 1.6.6 or higher.
Refer also to the OUTPut:DISABLEPHase command.
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<b>
Example
OUTP:ZERO 1
Query Format
OUTPut:ZEROprogram?
Returned Data Format
<b>
Query Example
OUTP:ZERO?
1