AFX SERIES® OPERATION MANUAL
SECTION 7:
Rear Panel, Connectors and Protection
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 172 of 402
7.3.5
Dedicated Function Digital Inputs
Remote Inhibit
External Trigger Input
External Phase Sync Input
7.3.5.1
Remote Inhibit
REMOTE
INHIBIT
The mode can be selected from the SYST (SYSTEM) menu screen using the Interfaces,
Remote Inhibit selection as shown below.
There are three modes of operation for this input:
4.
Disable mode: In his mode, the Remote Inhibit inputs are not active so this function
is turned off. No short between the RI pins is required to enable the output in this
mode.
5.
Remote Inhibit mode: The two pins have to be shorted for the output to be enabled
from the front panel or remote command. This is a necessary but not a sufficient