AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 231 of 402
Command Syntax
PROGram:TRANsient:CSC
Description
This command turns the CSC Mode during transient execution on (1)
or off (0). This allows load regulation adjustment during transient
execution. However, keep in mind that CSC adjustments are based on
RMS voltage measurement, which take several cycles to run so
turning on CSC on fast changing voltage transients is not
recommended as the CSC may interfere with the programmed
transient voltages.
Note 1: In UPC compatibility mode, the CSC is always off during
transient execution regardless of this setting to match the UPC
controller operation.
Note 2: This command is available on units with firmware revision
1.3.0 or higher.
Parameters
< 0 | OFF | 1 | ON >
Parameter Format
<cr>
Example
PROG:TRAN:CSC ON
Query Format
PROGram:TRANsient:CSC?
Returned Data Format
<cr>
Query Example
PROG:TRAN:CSC?
1
Command Syntax
PROGram:TRANsient:FROM
Description
This command sets the first transient step number at which step
mode execution is to begin. The step specified must exist or an error
message will be generated.
Parameters
STEP number from 1 ~ 99
Parameter Format
<nr1>
Example
PROG:TRAN:FROM 5
Query Format
PROGram:TRANsient:FROM?
Returned Data Format
<nr1>
Query Example
PROG:TRAN:FROM?
5
Command Syntax
PROGram:TRANsient:TO
Description
This command sets the transient step number at which step mode
execution is to end. The step specified must exist or an error message
will be generated.
Parameters
STEP number from 1 ~ 99
Parameter Format
<nr1>
Example
PROG:TRAN:TO 25
Query Format
PROGram:TRANsient:TO?
Returned Data Format
<nr1>
Query Example
PROG:TRAN:TO?
25