700 NDIR Analyzer
6-20-2017
160
ASTF: Error Status
Command
Response
Description
_ASTF_K0
_ASTF_s_f1_f2_f3_...f32
Current error numbers are responded.
1: Check flow channel 1
2: Check flow channel2
3: Check flow channel 3
4: Check external analog 1
5: Check external analog 2
6: Check barometer
7: Check analyzer temperature
8: Channel 1 is not calibrated
9: Channel 2 is not calibrated
10: Channel 3 is not calibrated
11: Channel 1 concentration 1 warning
12: Channel 2 concentration 1 warning
13: Channel 3 concentration 1 warning
14: Channel 1 concentration 2 warning
15: Channel 2 concentration 2 warning
16 : Channel 3 concentration 2 warning
17: Check channel 1 temperature
18: Check channel 2 temperature
19: Check channel 3 temperature
20: Check channel 1 EPC voltage
21: Check channel 2 EPC voltage
22: Check channel 3 EPC voltage
23: Channel 1 range overflow
24: Channel 2 range overflow
25: Channel 3 range overflow
26: Channel 1 ADC range overflow
27: Channel 2 ADC range overflow
28: Channel 3 ADC range overflow
29: Channel 1 ADC range underflow
30: Channel 2 ADC range underflow