700 NDIR Analyzer
6-20-2017
73
Span Gas Deviations Verifying
→
→
→
→
→
Press
from the Auto Cal Deviations menu to view the Verifying Span Deviations
screen.
Use the left or right arrow keys
to select the component you wish to view.
Measured Value
is the averaged concentration during the Verifying Span stage of
sequenced and auto calibrations.
Variance
is the difference of the measured value and span gas concentration.
% FS
is the percent of full scale related to the calibrated range limit.