700 NDIR Analyzer
6-20-2017
146
AK Protocol
Data Description
Each command begins with STX (Start of Text) in the first byte. The “don’t care” byte can be
any ASCII character. Generally, a blank space or an underscore ( _ ) is used to increase
readability. The four function bytes represent the AK command. A blank space comes next,
followed by K and the channel number. The analyzer is a single-channel device, and because of
that, the channel number is usually 0. For delimiting the command parameters from the channel
number, another blank follows. This may be followed by command parameters with variable
lengths. Every command ends with the ETX (End of Text) character. The Error Status byte does
not indicate the real number of errors. For Error Status, use the ASTF command.
Example:
Using Windows
®
HyperTerminal for Serial RS232 Communications with CAI 700
Series Analyzers requires:
1.
HyperTerminal Software
2.
Windows PC/Laptop
3.
Null Modem Cable
Setup Procedure: run HyperTerminal and configure as shown:
Other similar simple terminal programs should allow similar settings.
Summary of Contents for 700 NDIR
Page 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Page 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Page 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Page 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Page 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...