700 NDIR Analyzer
6-20-2017
157
AKAK: Calibration Gas Concentrations
Command
Response
Description
_AKAK_Km
_AKAK_s_M1_w.w
_M2_x.x
_M3_y.y
_M4_z.z
All existing calibration gas values for channel m.
_AKAK_Km_Mn
_AKAK_s_Mn_z.z
Calibration gas value of channel m range n.
AKAL: Percent Deviations of Last Accepted Calibration
Command
Response
Description
_AKAL_Km_
_AKAL_s_M1_z.z_y.y_x.x_w.w
_M2_z.z_y.y_x.x_w.w
_M3_z.z_y.y_x.x_w.w
_M4_z.z_y.y_x.x_w.w
Percent Deviation of channel m ranges M1 to
M4.
z.z: Zero gas relative to last calibration
y.y: Zero gas absolute to factory calibration
x.x: Span gas relative to last calibration
w.w: Span gas absolute to factory calibration
AKEN: Device Identification
Command
Response
Description
_AKEN_K0
_AKEN_K1
_AKEN_K2
_AKEN_K3
_AKEN_s_devicename
_AKEN_s_model
_AKEN_s_serial number
_AKEN_s_sample pressure
Device identification is responded.
Device Model.
Device Serial Number.
Suggested sample input pressure.