700 NDIR Analyzer
6-20-2017
168
Configuration Commands
ECO2: Set CO2 Correction Factor s
Command
Response
Description
_ECO2_Km_z.z_y.y_x.x_w.w
_ECO2_s
Set CO2 correction factors for channel m.
z.z = Offset
y.y = Minimum analog input
x.x = 1
st
order coefficient
w.w = 2
nd
order coefficient
EDAL: Set Diagnostic Alarm Limits
Command
Response
Description
_EDAL_K0_ai.min_
a1.max_..._a16max
_EDAL_s
Set all alarm limits.
1: Channel 1 Flow
2: Channel 2 Flow
3: Channel 3 Flow
4: External analog 1
5: External analog 2
6: Barometer
7: Analyzer temperature
8: Channel 1 Concentration 1 / Concentration 2
9: Channel 2 Concentration 1 / Concentration 2
10: Channel 3 Concentration 1 / Concentration 2
Summary of Contents for 700 NDIR
Page 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Page 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Page 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Page 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Page 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...