700 NDIR Analyzer
6-20-2017
159
APAR: Auto Calibration Tolerance Values
Command
Response
Description
_APAR_Km_SATK
_APAR_s_z.z_y.y_x.x_w.w
Auto calibration tolerance value (%) for channel
m.
z.z: Range 1
y.y: Range 2
x.x: Range 3
w.w: Range 4
ARAW: Raw Detector Volts
Command
Response
Description
_ARAW_K0
_ARAW_s_z.z_y.y_x.x_t
Raw Detector Volts for all channels.
z.z : Channel 1
y.y : Channel 2
x.x : Channel 3
t = Timestamp (1/10 sec.)
_ARAW_Km
_ARAW_s_z.z_t
Raw Detector Volts of channel m.
t = Timestamp (1/10 sec.)
ARMU: Raw Engineering Value
Command
Response
Description
_ARMU_K0
_ARMU_s_z.z_y.y_x.x_t
Raw Engineering Value before linearization,
offset and span correction for all channels.
These are the values used to calculate the
polynomial coefficients.
t = Timestamp (1/10 sec.).
_ARMU_Km
_ARMU_s_z.z_t
Raw Engineering Value before linearization,
offset and span correction for Channel m.
These are the values used to calculate the
polynomial coefficients.
t = Timestamp (1/10 sec.).