Calibration Setup
→
→
→
The Calibration Setup menu provides all the parameters necessary for completing a
successful calibration.
To access the Calibration Setup menu, select
from the
Calibration menu.
All parameters on the Calibration Setup menu apply to Automated Calibration. The
following also apply to Manual Calibration: Cal Pump/Valves, Auto Cal/Check,
Deviation Limits and Analog Hold on Cal. All settings should be verified for correct
information before a manual or automated calibration is attempted.
Please note that the Calibration Setup menu shows the current settings on the right side of
the screen after the ellipsis
(…). Example:
Cal Pump/Valves . . . Valves.
Auto Calibration Schedule
→
→
→
→