Reset Offsets and Gains
→
→
→
→
Pressing
from the Reset Factory Settings menu will prompt the operator to confirm
resetting Offsets and Gains for all ranges of the current channel. Use the left or right
arrow keys
to select the component you wish to reset. Pressing
(Yes)
from this screen resets the Offset and Gain factors to factory default settings (0 and 1
respectively) and brings you to this confirmation screen:
Offset and Gain factors are created when the analyzer is zeroed and spanned.
If the Offsets and Gains are reset, the analyzer must be zeroed and spanned again
before use.
All recorded deviations will be set to zero.