700 NDIR Analyzer
6-20-2017
149
Acknowledgement Command
Byte
Character
Explanation
1
st
Byte
STX
ASCII Code 02
2
nd
Byte
Don’t Care
Any ASCII code
3
rd
Byte
Function Code 1
Echo of the AK Instruction Command
4
th
Byte
Function Code 2
Echo of the AK Instruction Command
5
th
Byte
Function Code 3
Echo of the AK Instruction Command
6
th
Byte
Function Code 4
Echo of the AK Instruction Command
7
th
Byte
Blank
8
th
Byte
K
9
th
Byte
0
10
th
Byte
Blank
D
AK acknowledgement parameters; lengths variable
A
AK acknowledgement parameters; lengths variable
T
AK acknowledgement parameters; lengths variable
A
AK acknowledgement parameters; lengths variable
nth Byte
ETX
ASCII Code 03
Example:
<STX>
02H Example: STZ 0 SREM STBY
Don’t Care
Adjustable, factory setting 20H
Function Code
Same code as command package (e.g. ASTZ)
Space
20H
Status
0 without error or 1 to 9 when error (see also ASTF command)
Space
20H (only if followed by data, otherwise <ETX>)
Data
Parameter (depending on the command)
<ETX>
03H