700 NDIR Analyzer
6-20-2017
81
Range Limits
→
→
→
or
→
The analyzer is factory-configured with four physical ranges (1 - 4).
The operator can
change the number of ranges and select a specific full-scale concentration in ppm. From
the Range Setup menu press
to access the Range Limits screen.
Use the left or right arrow keys
to select the component you wish to view or
change. To change the Range Limits from the factory settings, use the up or down arrows
to move the highlight to the field you intend to change. Press
to open the
field to change the value in ppm. Press
again to close the field. Press
to save
your changes. To initiate the saved changes, press
, then press
and select new
ranges.
NOTES:
1.
If Ranges are factory linearized, the range limit must not exceed linearized factory
range.
2.
The Range Limit values must be set in ascending order.