700 NDIR Analyzer
6-20-2017
158
AKON: Measured Concentration Value
Command
Response
Description
_AKON_K0
_AKON_s_z.z_y.y_x.x_t
Current Measured Value of all channels.
z.z : Channel 1
y.y : Channel 2
x.x : Channel 3
t = Timestamp (1/10 sec.).
_AKON_Km
_AKON_s_z.z_t
z.z : current Measured Value of channel m.
t = Timestamp (1/10 sec.).
AMBE: Measuring Range Limit
Command
Response
Description
_AMBE_Km
_AEMB_s_M1_w.w
_M2_x.x
_M3_y.y
_M4_z.z
All existing measuring range limits of channel m.
_AMBE_Km_Mn
_AEMB_s_Mn_z.z
Measuring range limit of channel m range n
AMBU: Upper and Lower Range Switchover Values for Auto Range
Command
Response
Description
_AMBU_Km
_AMBU_s_M1_w.w_W.W
_M2_x.x_X.X
_M3_y.y_Y.Y
_M4_z.z_Z.Z
Lower and upper range switchover values for
auto range of channel m.
_AMBU_Km_Mn
_AMBU_s_M1_w.w_W.W
Lower and upper range switchover values for
auto range of channel m range n.