700 NDIR Analyzer
6-20-2017
155
ADRU: Pressures and Electronic Pressure Control Valve Voltage
Command
Response
Description
_ADRU_K0
_ADRU_s_z.z_y.y_x.x_w.w
_Y.Y_X.X_W.W
Pressures and EPC valve voltages.
z.z: Environment pressure.
y.y: sample pressure channel 1
x.x: sample pressure channel 2
w.w: sample pressure channel 3
Y.Y: % of Sample EPC Volts for channel 1
X.X: % of Sample EPC Volts for channel 2
W.W: % of Sample EPC Volts for channel 3
_ADRU_Km
_ADRU_s_ % voltage
% of Sample EPC Volts for channel m
ADUF: Flows
Command
Response
Description
_ADUF_K0
_ADUF_s_z.z_y.y_x.x
Sample gas flow of all channels.
_ADUF_Km
_ADUF_s_z.z
Sample gas flow of channel m.
AEMB: Get Measuring Range
Command
Response
Description
_AEMB_K0
_AEMB_s_Mn_Mn_Mn
Current measuring range of all channels.
_AEMB_Km
_AEMB_s_Mn
Current measuring range of channel m.
AENT: Query Calibration Gas Flow Settings (Pumps or Valves)
Command
Response
Description
_AENT_K0
_AENT_s_x
Current calibration gas flow setting.
x: 10 = Pump
11 = Valves