700 CLD / O2
5-04-2017
99
NO/NO
x
/NO
2
Mode Times
→
→
→
→
The NO/NO
x
/NO
2
Mode Times screen allows the operator to set the Purge Time and
the Integration Time.
From Measure Settings menu press
to access the
NO/NO
x
/NO
2
Mode Times screen.
The Purge and Integration times are set to allow adequate purging and integration times
between the NO and NO
x
cycles. Any gas remaining in the sample stream from the
previous mode is purged before integration into the next mode. All NO and NO
x
readings
are displayed as averaged values, and the sample read times can be adjusted by the
operator. See
NO/NO
x
/NO
2
Mode
for more information.
Use the Up/Down arrows
to highlight the field you intend to change. Press
to open the field to change the time (in seconds). After making your change, press
again to close the field. Press
to exit the screen and return to the Measure
Settings screen.