750 Manual
5-04-2017
157
AKAK: Calibration Gas Concentrations
Command
Response
Description
_AKAK_K0
_AKAK_s_M1_w.w
_M2_x.x
_M3_y.y
_M4_z.z
All existing calibration gas values are responded
_AKAK_K0_Mn
_AKAK_s_Mn_z.z
Calibration gas value of Range n
_AKAK_K1
_AKAK_s_M1_w.w
K1 is for O
2
option
AKAL: Percent Deviations of Last Accepted Calibration
Command
Response
Description
_AKAL_K0_
_AKAL_s_M1_z.z_y.y_x.x_w.w
_M2_z.z_y.y_x.x_w.w
_M3_z.z_y.y_x.x_w.w
_M4_z.z_y.y_x.x_w.w
_02_z.z_y.y_x.x_w.w
Percent Deviation of ranges M1 to M4 and O
2
with O
2
option
z.z: Zero gas relative to last calibration
y.y: Zero gas absolute to factory calibration
x.x: Span gas relative to last calibration
w.w: Span gas absolute to factory calibration
AKEN: Device Identification
Command
Response
Description
_AKEN_K0
_AKEN_K1
_AKEN_K2
_AKEN_K3
_AKEN_K4
_AKEN_s_devicename
_AKEN_s_model
_AKEN_s_serial no
_AKEN_s_Air pressure
_AKEN_s_Sample pressure
Device identification is responded
Device model
Device serial number
Air pressure
Sample pressure